Features:

  • Closed loop HEPA-filtration 
  • Dual Sterilization Cycles
  • Coved interior corners
  • CuVerro Antimicrobial Surfaces
  • Temperature Uniformity
  • CO2 Gas Accuracy
  • Growth Condition Recovery

 

 

Features:

Standard Features:

  • NuTouch Electronic Control System
  • 100% Stainless Steel Coved Interior Chamber 
  • Dual Temperature Sensor 
  • Infrared (IR) CO2 Sensor 
  • Dual Sterilization Cycles (NU-5710, 5720, 5731, 5741)
  • Humidity Control System (NU-5720, 5741)
  • O2 Control System (NU-5731,5741)
  • Four Stainless Steel Shelves
  • Eight Stainless Steel Shelf Guides 
  • Four Wall Brackets 
  • Right Hinged Door Swing 
  • Remote Alarm (output contacts)
  • 4 to 20 mA Analog Output RS-485 Communication USB-port
  • Co2 Sample Port 
  • Adjustable Leg Levelers
  • Access Port and Plug with Breather Holes 
  • One Water Pan
  • One Electrical Cord 2 m

Optional Features: 

  • CuVerra Antimicrobal Copper Surface (Interior Chamber)
  • CuVerra Antimicrobal Copper Surface (Shelving and Guide Brackets)
  • Automatic CO2 Tank Switch (External)
  • Left Hinged Door Swing 
  • Additional Stainless Steel Shelves with Guide Brackets 
  • CO2 analyzer Fyrite Kit (dry 0-20%)
  • Replacement Fluid for CO2 Analyzer 
  • CO2 Regulator (2 Stage)
  • Platform with Castors 

Temperatuur Control System:

  • Temperature Sensor Type: Precision Integrated Circuit
  • Default Set Point: 37°C
  • Chamber temerature Range: 5°C to 55°C  (5°C Above Ambient to 30°C Max. Ambient)
  • Chamber Temerature Uniformity: plus minus 0,3 °C @ 37°C.
  • Temperature accuracy: plus minus 0,1°C
  • Temperature recovery: 0,12°C / Minute Average 
  • Temperature Display Resolution: 0,1°C
  • Minimum Qualifications for Sterilization:       145 DEG Cycle 135 °C

                                                                     95 DEG Cycle 85 °C

Electrical Requirements:

  • Startup Powerl: 625 watt
  • Running Power: 250 watt, 60 Hz
  • Decon cycle: 995 watt
  • Heat Rejected: 14 BTU/min

Utility Connections:

  • Gas Connections: 0,25 in, (6,3 mm) Tubing Connections
  • Gas Input Pressure: 20 PSIG (1,4 BAR) Input Pressures Maximum, Two-Stage Gas Regulators Required.

Co2 Control Systems:

  • Co2 Sensor Type: Infrared Single Source Dual Wave Length  
  • Co2 lControl Logic: Fixed Algorithm / Manual Environmental Adaptable.
  • Default Set Point: 5%
  • Co2 Range: 0,1 to 20%
  • Co2 Accuracy: plus minus 0,1%
  • Co2 Recovery: Up to 5%, -0,50% / +0,20% in 5 Minutes Average. 
  • Co2 Display Resolution: 0,1%

 

RH (NU-5720 / 5741)

  • Default Set-Point: 90%
  • RH Range: 5% Above Ambient to 90%
  • RH Accuracy: plus minus 3%
  • RH Recovery: 90% +5% / -3% 25 min.

O2 (NU-5731 / 5741)

  • Zirconia Ceramic Sensor
  • Default Set-Point: 21%
  • O2 Range: 0.5 to 21%
  • O2 Accuracy: ± 0.25%
  • O2 Recovery: 5% ± 2% / 20 min

 